Lithium ion selective electrode (ISE)
Measure ISE Lithium ion in aqueous based solutions.
NT Sensors, manufacturer of electrochemical sensors since 2008 has developed its own lithium ion selective electrode.
With good selectiviy towards other cations, Lithium ISE can be used in different applications where lithium levels require to be determined.
Good response time <60 sec. Nernstian Sensitivity. Dry-storage. Small sample volumes required.
...
SPECIFICATIONS
Range (concentration): 0,09 to 9000 mg/L / 5×10-6 to 0,5 (mol/L)
Reproducibility (precision): ±10% of full scale (calibrated 10 to 1000 mg/L)
Interfering ions: K+, Na+, H+
pH range: 2 to 12 (no pH compensation)
Temperature range: 4 – 40°C (no temperature compensation)
Electrode slope: 54 ±5 mV
Electrode resistance: 1–4 MΩ
Immersion minimum: 2 mm (without the protective cap)
Body diameter: see different sensors

Mini ION Lithium
Single Probe
BNC or USB
Nano 2.5: Miniature electrode for Lithium
Nano probes that are ideal for research applications or ultra-low volume purposes. Capable to work in al positions (vertical, horizontal, upside down). In addition, the carbon nanotubes provides the real miniaturization. Moreover, retaining good performance in terms of ease of use and robustness of operation.
REF: M007
Diameter: 2,5 mm
Minimum sample volume: 0,25 ml
More information about
Nano 2.5
–View product–