Dual Nitrite ISE + pH Meter This nitrite ion meter (laboratory precision) is a nanotechnology instrument used to obtain reliable data using data combining a nitrite selective electrode and a pH electrode. Multiparameter instrument for analysis of NO2- ion concentrations and pH ( imacimus 2The nitrite electrode meter is used for simultaneous measurement of nitrite ions and pH in aqueous solutions.
Examples of where to apply nitrite control:
The ion analyser for nitrite and pH is supplied with powerful analysis software that displays the measurement (and allows identification) and allows customised calibration up to 5 customised standard points, among other features. Continuous monitoring is also available for the study of short-term reactions or processes.
The nitrite ion sensor is a replacement module and can be exchanged for a new nitrite sensor or any other sensor available in our catalogue.
Note the limited lifetime of the ISE sensor, even if they are stored on a shelf or installed in a probe. Different sensors differ by several months under the same conditions.
‘]
pH IÓNICO único
Cable USB
NT Sensors software
– 1x dual channel ION / mV / pH / NITRITE METER for use with PC / laptop.
Ref.
Imacimus 2 Meter – 1x Software for Windows (free download) – 1x USB data/power cable – 1x SINGLE ION PROBE.
Support for up to 1 Mini ISE.
– 1x Mini nitrite electrode – 1x Reference/pH electrode – 1x Calibration Standard 10 mg/L Nitrite 100mL – 1x Calibration Standard 100 mg/L Nitrite 100mL – 1x Calibration Standard 1000 mg/L Nitrite 100mL – 1x Buffers pH4 (50 ml) – 1x Buffers pH7 (50 ml) – 1x User’s Manual
Your pH and nitrite ion tool kit comes with all the elements you need to start reading samples.
– 1x dual channel ION / mV / pH / NITRITE METER for use with PC / laptop.
Ref.
Imacimus 2 Meter – 1x Software for Windows (free download) – 1x USB data/power cable – 1x SINGLE ION PROBE.
Support for up to 1 Mini ISE.
– 1x Mini nitrite electrode – 1x Reference/pH electrode – 1x Calibration Standard 10 mg/L Nitrite 100mL – 1x Calibration Standard 100 mg/L Nitrite 100mL – 1x Calibration Standard 1000 mg/L Nitrite 100mL – 1x Buffers pH4 (50 ml) – 1x Buffers pH7 (50 ml) – 1x User’s Manual
Your pH and nitrite ion tool kit comes with all the elements you need to start reading samples.
Employing the same operating principle as standard laboratory electrodes
The potentiometry technique (ISE) is relatively inexpensive and easy to operate.
Ion selective electrodes have a wide concentration measurement range.
Specifications
Data
ISE range
0.4 to 4,000 mg/L
ISE/pH resolution
0 to 14 pH is 0.01 pH
0 to 1.0 ppm is 0.1 ppm
1 to 9,999 ppm is 1 ppm
ISE accuracy
±4 % or 8 % * In the absence of interfering ions, the accuracy may be better than ±1 mV, i.e. corresponding to ±4 % of the measurement.
(For divalent ions, ±2 mV; ± 8 %)
Repeatability ISE
±10% monovalent ±20% divalent ** Repeatability in the measurement of a standard solution after calibration.
ISE/pH calibration
From 2 to 5 points.
Fully customizable calibration standards.