Technical questions
1) What do CNT Ion Selective Electrode measure?
2) What is the difference between activity and concentration?
3) What is ISAB and when is it used?
4) What is meant by Interference in CNT_ISE measurements?
5) What is new in NT Sensors reference electrode?
6) Do I need to replace or top-up any electrode filling solutions?
7) What should I do if the measured electrode slope is outside the specified limits?
8) After immersing the electrodes, how long should I wait before taking a reading?
9) What precision can I expect from a CNT_ISE measurement?
10) What is meant by "drift" in ISE measurements?
11) What is the effect of temperature change on CNT ISE measurements?

12) How frequently should I re-calibrate the electrodes?

13) What is the concentration range and detection limit for a CNT_ISE measurement?

14) How should I store the electrodes between analytical sessions?

15) What is the shelf life and operating life of a CNT_ISE?

16) Does the operating life of a CNT_ISE depend on the number of samples analysed?

17) Should I worry if the measured voltage is negative?
18) What are the advantages of using potentiometric measures?

1) What do CNT Ion Selective Electrode measure?
CNT ISEs measure the activity of free ions in aqueous solutions.

2) What is the difference between activity and concentration?
Activity is the effective concentration - i.e. that portion of the ions which are free to take part in a given reaction - in this case coming into contact with the membrane surface. Activity is always less than concentration due to inter-ionic interactions in the solution, which inhibit the movement of ions and prevent some of them from reaching the membrane. It becomes proportionately less as the concentration increases. In practice this effect is negligible (within the error limits of the measurement) below about 0.01M for monovalent ions and 0.001M for divalent ions. The difference between activity and concentration is expressed as the Activity Coefficient.

3) What is ISAB and when is it used?
Ionic Strength Adjustment Buffer is added equally to samples and standards to minimise any errors due to differences in ionic strength between samples and standards which will cause differences in activity coefficents which can cause the concentration to be understimated by up to 50 or 60% in the worst cases. In some cases ISABs can also include ingredients which minimise interference effects, and ensure that the pH is optimum for the CNT_ISE measurement.

Furthermore, for some ions, the addition of ISAB can help to reduce the time required to reach a stable reading after immersing the electrodes in a new solution.
The addition of ISAB is not needed for the CNT_ISEs.For highly accuracy measures, when the uncertainty required must be very low, the activity of the ion must be taken into account, as factor of the ionic strenght.

4) What is meant by Interference in CNT_ISE measurements?
Ion-selective electrodes are not completely ion-specific. All are sensitive to some other ions to some extent. The degree of sensitivity to another ion is given by the Selectivity Coefficient.

5) What is new in NT Sensors reference electrode?
CNT_ISE Multi Ion and CNT_ISE Combi Ion include NT Sensors reference electrode, CNT_RE. It is an innovative reference electrode, which combines the classic properties of a reference electrode with new technologies. CNT_RE is compatible with all CNT_ISEs.

6) Do I need to replace or top-up any electrode filling solutions?
No. All CNT_ISEs are all-solid-state and contain no liquid or gel solutions - so there is nothing to replace.
The only need for CNT ISEs is to keep the protector cap in the sensor to prevent faster aging.

7) What should I do if the measured electrode slope is outside the specified limits?
If the slope is only a few millivolts outside specification but is stable and reproducible then the CNT_ISEs can still be used satisfactorily, although the lower the slope the higher the errors on the measurement of activity (or concentration).
The aging of part of the samples will cause a progressive decrease in the CNT ISE slope. It could be also due to membrane contamination. In this case it has to be conditioned at least for 8 hours in a solution of the target ion (0.1M / 1000 mg/L)

8) After immersing the electrodes, how long should I wait before taking a reading?
Most electrode systems require about two minutes to reach a completely stable reading. Nevertheless, CNT_ISE get to 90% of the final value in less than thirty seconds, so it depends on your precision requirements as to whether you wait for complete stabilisation or not. 

9) What precision can I expect from a CNT_ISE measurement?
Assuming that there are no systematic errors from interference or activity coefficient effects, the precision depends on the error in the measurement of the electrode potential (mV) and the slope of the calibration line:- a 1mV error represents a ~4% error in the concentration for monovalent ions (slope~54) and an ~8% error for divalent ions (slope~27), when working in the normal linear range of the electrodes.
Nevertheless, it has been shown that, under optimum conditions, a precision of better than ± 2% can be achieved by making frequent recalibrations and taking multiple readings of the electrode signals.

10) What is meant by drift in ISE measurements?
Drift is the gradual change in electrode response over a period of time. The rate and extent of drift can vary depending on which particular CNT_ISE is being used and the age and degree of contamination of the electrodes. If a series of samples are measured repeatedly it will be observed that each successive reading of the same solution will be slightly different from the previous one. This classic problem in potentiometric measures can be overcome by frequently measuring one calibration solution in between sample measurements, then re-calibrating when the mV value has drifted beyond an acceptable level - depending on precision requirements. 

11) What is the effect of temperature change on CNT_ISE measurements?
CNT ISE response depends on the temperature. Thus, to avoid any errors due to temperature change it is advisable that standards were at the same temperature as the sample solutions. If the sample temperature deviates by more than about 2°C from the original calibration temperature, it is advisable to recalibrate.

12) How frequently should I re-calibrate the electrodes?
It depends on the precision requirements for the results and the rate of drift of the electrode system. If only an order of magnitude measurement is required then it may only be necessary to calibrate once a day or even less frequently. Apart from drift, any temperature changes (greater than 2°C) will cause a change in the calibration. For the most precise results, it may even be beneficial to calibrate between every sample. 

13) What is the concentration range and detection limit for a CNT_ISE measurement ?
These figures are given in the specifications for each CNT_ISE (you can find them here) and they are very similar to classical potentiometric methods.
In general it can be noted that most ISEs cannot be used for concentrations below about 0.1 ppm and many will only give a linear response and reasonable precision above 1 or 2 ppm. The upper concentration limit is usually quoted as 1Molar, but in practice it is difficult to obtain reliable results above about 0.5Molar because of uncertainties in the effect of high ionic strength on the activity coefficient.

14) How should I store the electrodes between analytical sessions?
In general, CNT_ISEs should always be rinsed with de-ionised water and gently dabbed dry with a low-lint tissue. They should not be left more than a few seconds in de-ionised water. Finally, the protective cap should be replaced to protect the membrane from atmospheric oxidation/corrosion.

15) What is the shelf life and operating life of a CNT_ISE?
CNT_ISE are unusually robust and durable. The shelf life of a new unused solid-state CNT_ISE is many years (providing that the protective cap is kept in place to limit any atmospheric corrosion). They are guaranteed for six months during normal operation, but in practice have been found to last much longer than this.

16) Does the operating life of a CNT_ISE depend on the number of samples analysed?
Experience has shown that electrodes can be immersed continuously for months without deteriorating. This demonstrates that, for individual sample measurements with average immersion times of two or three minutes, many thousands of samples can be measured with one electrode.
CNT_ISE operation is based on ion-exchange / charge-transfer principles. Thus there is any consumption of materials or wear and tear on components when samples are analysed. Assuming that the electrode is not exposed to damaging chemicals, the only deterioration is due to the gradual corrosion (oxidation, hydration, leaching) of the membrane, the plastic body, and the adhesive used to attach the membrane to the body (or to physical damage to the membrane surface or to the pin connecting to the electrode head). The extent of the corrosion is affected by the length of time the electrode is immersed or exposed to the atmosphere rather than the number of samples analysed.

17) Should I worry if the measured voltage is negative?
It is not necessarily a problem. The measurement range of a CNT_ISE can be positive or negative, it depends on some elements.

18) What are the advantages of using potentiometric measures?
It is a fast and efficient method to measure ions. This method is economical, easy handled and it has a wide range of applications and continued monotoring.
It is used in laboratories, industries, environmental, agriculture...Furthermore, it doesn't require large volums of sample or reactives and it is a non-destructive testing. It is possible to measure coloured solutions or with presence of solid particles.


 
© NT Sensors 2011